VISCOSITY TEST OF BITUMEN (IS 1206-PART-2)
|沥青的绝对粘度试验
Reference
IS: 1206 (Part 2)– Method for Testing Tar and Bitumen Materials: Determination of Absolute Viscosity
Test Equipment
The following testing equipment is needed to measure absolute viscosity of viscosity graded paving bitumen (是:73-2006) at 60 C in accordance with IS:1206 (Part 2) (similar toASTM D 2171), which uses a vacuum capillary viscometer. It should be noted that any other viscometer such as rotational viscometer cannot be used to measure the absolute viscosity at 60 C since the bitumen in non Newtonian in nature at this temperature.
While obtaining quotations specify the following requirements. Although Asphalt Institute vacuum viscometer and Modified Koppers vacuum viscometer can also be used, the following refers to Cannon-Manning vacuum viscometer only.
Complete Absolute Viscosity Testing Equipment conforming to IS:1206 (Part 2) Method for Testing Tar and Bitumen Materials: Determination of Absolute Viscosity with the following components:
- Constant Temperature Bath– A suitable bath for immersion of at least 6 vacuum capillary viscometer tubes with a digital temperature controller. The accuracy of the temperature in the bath should be +/- 0.1 C throughout the bath.
- Silicone Bath Oilsuitable up to 150 C
- Vacuum System– Capable of maintaining a vacuum within +/- 0.05 cm of the desired level up to and including 30 cm of mercury. The system shall consist of vacuum pump, moisture trap, vacuum regulator, manometer with electronic controller, bleed valve, all interconnecting tubing/piping, and any other accessories as needed to complete the vacuum system.
- Thermometer for Bath– Mercury in glass, range 37.8 to 82 C, and graduations of 0.2 C.
- Timing Device–A stop watchor stop clock capable of reading up to ½ second.
- 炮弹 - 曼宁真空粘度仪– With manufacturers’ calibration certificate, viscometer holder and silicone cork. Size 12 and Size 13 (one each) [Size 12 is suitable for testing VG-10 and Size 13 is suitable for testing VG-20, VG-30, and VG-40 bitumen.]
- Viscometer Stand– for holding 6 viscometers.
- Installation of the quoted instrument in client’s laboratory and operational training.
Note:Only theCannon-Manning viscometertubes need to be imported at this time, the remaining equipment such as bath and vacuum system are available in India.
However, be aware of water bath and vacuum control systems manufactured in India, many do not give the accuracy and consistency needed for the test (viscosity is very sensitive to small variations in temperature and vacuum level). Try them out carefully before buying them. If not satisfactory, these systems may have to be imported.
Test Procedure
Preparation of Sample:
- 热的沥青样品温度than 90 C above its approx. softening point until is has become sufficiently fluid (like motor oil) to pour easily (Pic-1)。
- Transfer about 20 ml into a suitable container (Pic-2) and maintain at a temperature of 135 +/- 5.5 C stirring occasionally to allow entrapped air to escape.
- Pour the hot bitumen in the Canning-Manning vacuum viscometer through the larger diameter filling tube A so that bitumen is within +/- 2 mm of the fill line E. Place the charged viscometer in an oven or bath maintained at 135 +/- 5.5 C for a period of 10 +/- 2 minutes to allow larger air bubbles to escape.
Testing:
- Maintain the test bath temperature at 60 +/- 0.1 C. Place the charged viscometer vertically in the test bath with the help of a holder so that that the uppermost timing mark is at least 2 cm below the surface of the bath liquid (图三)。
- Establish a vacuum of 30 +/- 0.05 cm of mercury in the vacuum system and connect to the viscometer with the valve closed. After the viscometer has been in the bath for 30 +/- 5 min, open the valve and allow the bitumen to flow in the viscometer.
- Measure the time required (to within +/- 0.5 sec) for the leading edge of the meniscus to pass between successive pairs of timing marks (Pic-4 and 5)。
- Report the first flow time which exceeds 60 sec between a pair of timing marks, noting the identification of the pair of the timing marks.
计算:
Calculate (Pic-6) and report the absolute viscosity in poises to three significant figures as follows:
Viscosity in poises = K t
Where:
K = Calibration factor in poise per second supplied with the viscometer tube for the pair of timing marks where the flow time exceeded 60 seconds.
t = flow time in seconds
Report:
Report the test temperature and vacuum level with the viscosity test results such as viscosity in poises @ 60 C and 300 mm mercury vacuum.
Note:Kinematic viscosity at 135 C also needs to be determined for the paving bitumen as specified inIS: 73-2006. At this temperature, bitumen flows readily so no vacuum needs to be applied. Follow IS: 1206 (Part III) Methods for Testing Tar and Bituminous Materials: Determination of Kinematic Viscosity, to select the proper viscometer tube given in there. The same oil bath as used for absolute viscosity can be used, just the viscometer tube will be different (Pic-7)。作为替代,人们还可以使用旋转粘度计,这也可以提供相同的结果,因为沥青是牛顿本质上的135℃。鼓励购买和使用旋转粘度计,因为它也可用于改性粘合剂,也可以用于改性粘合剂SuperPave性能分级(PG)粘合剂,预计将在不久的将来采用。同样,旋转粘度计不能用于测量60℃的绝对粘度。
Article Written By
Prof. Prithvi Singh Kandhal
斋浦尔,印度
there are no +- limits are give in IS code for cannon manning viscometer tube. So is there any other source from where we could get the limitations for its dimensions?
IS IT NECESSARY TO CHECK BOTH ABSOLUTE VISCOSITY AND KINEMATIC VISCOSITY?
,IF YES WHAT IS THEIR IMPORTANCE ?,,,,,,,,,AND WHY THEY ARE NOT CONDUCTING AT SAME TEMPERATURE,,,,,,,?
Sir, Which no of tube use for kinematic viscosity of bitumen vg40?
It is good question, usually people mistaken the upper most third bulb, as this bulb is for safe gaurd for giving you time to close the vacuum system so that the bitumen should not flow to vacuum system. only two lower bulbs are to be accounted for.
Sir
Please rly me sir…
Sir
How to maintain bath temperature 135 along with Silicon fluid in 100 c capacity water bath.if we use silicon fluid how much of Qty need to take.
The silicon fluid required depends on the bath volume you are using, It is required that the liquid level in the bath should be kept about 25mm lower as when temperature rise the silicon fluid owing to high viscosity swells, Please select lowest viscosity oil (of about 50CsT) for least expansion on higher temperatures.
Sir
Asking in one bulb,we put cork after remove air bubble at 135 c sir
Small dought sir
在连接VISCUM泵的时,需要在相邻的侧面大道上拆下软木塞,如果需要删除它。
Make it solve sir full in confusion about that please
You should not use cork on any limb in Cannon Manning viscometer as if you use cork on any limb owing to vapour pressure generation at 135°C the rheology will automatically start in viscometer, for releasing air entrapped keep the viscometer at 135°C in an oven for 10 minutes and than put it in water/oil bath for 30 minute at a temperature of 60°C±0.1°C, apply vacuum and than take the time taken in filling in each bulb.
Sir, i have one concern in the above article. In the above the their is three bulb in the tube. which bulb time we used for calculating the absolute viscosity. or we can add the three of bulb viscosity and taking average of it.